Test Code LAB145 Amikacin, Random Plasma or Serum

Additional Codes

45050

Reporting Title

Amikacin Random

Methodology

Roche Diagnostics

Performing Laboratory

OhioHealth Laboratory Services-RMH Core Chemistry

Specimen Requirements

Submit only 1 of the following specimens:

 

Preferred:

Serum

Draw blood in a Gold-top serum gel tube(s. Spin down and send 1 mL of serum refrigerated. 

 

Note:  1. Date and time of draw are required on request form and specimen for processing.

2. A speckled-top serum gel tube(s) or a plain red-tube  is also acceptable.

3. Indicate serum 

Alternate:

Plasma

Draw blood in a mint green-top (heparin) gel tube(s). Spin down and send 1 mL refrigerated

 

Note:  1. Date and time of draw are required on request form and specimen for processing.

2. Indicate plasma on request form.

 

Specimen Stability Information

 

 

Specimen Type Temperature Time
Serum or plasma Refrigerated 48 hours capped
  Ambient  8 hours capped
  Frozen  4 weeks capped

 

 

Specimen Transport Temperature

Frozen/Ambient OK/Refrigerate OK

Reference Values

No established reference range for a random sample

Note:  Kanamycin has significant crossreactivity. If patient is receiving kanamycin, the amikacin test will be invalid.

Day(s) Test Set Up

Monday through Sunday; Continuously

.

Test Classification and CPT Coding

80150