Test Code LAB306 Factor VIII Assay, Blood or Plasma
Additional Codes
45585
Reporting Title
Factor VIII (8) Assay
Methodology
End point Clot Detection
Used to detect factor VIII deficiency.
Performing Laboratory
OhioHealth Laboratory Services-RMH Core Coagulation
Specimen Requirements
Submit only 1 of the following specimens:
Riverside Inpatients or Riverside Campus draw sites:
Testing must be performed within 4 hours of draw if submitting unprocessed whole blood.
Draw a full, light blue-top (citrate) tube; and send citrated whole blood. Invert several times to mix blood. Forward unprocessed whole blood promptly at ambient temperature only. Do not refrigerate.
Note:
1. This assay requires its own specimen.
2. Gross hemolysis, icterus, and lipemia may affect results.
3. Indicate whole blood on request form.
4. Label specimen appropriately (whole blood).
Outpatients/Outreach Clients:
Plasma-Frozen
Draw a full, light blue-top (citrate) tube. Spin down, remove plasma, spin plasma again, and place 1 mL of citrated platelet-poor plasma in plastic vial. (Glass vial is not acceptable.) Pour-off vial should be no more than 2/3 full. Send specimen frozen.
Note:
1. This assay requires its own specimen.
2. Double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious
results.
3. Gross hemolysis, icterus, and lipemia may affect results.
4. Indicate plasma on request form.
5. Label specimen appropriately (plasma).
NOTE: If multiple factor assays are requested, 2-3 (1ml) aliquots are required.
Specimen Transport Temperature
Ambient OK/Refrigerate NO/Frozen NO-Blood
Frozen OK/Refrigerate NO/Ambient NO-Plasma
Reference Values
51-154%
Critical value: <5%
Alerting category: First instance within 1 hour (initial result only, routine on results thereafter.
Day(s) Test Set Up
Monday through Friday (after 2 p.m. and on Saturdays and Sundays performed on emergency basis only)
Requests after 2pm on Saturday and Sunday require a Pathologist's approval.
Test Classification and CPT Coding
85240