Test Code LAB316 Heparin Anti-Xa Assay, Plasma
Additional Codes
47116
Reporting Title
Heparin Anti-Xa Assay, P
Methodology
Chromogenic Method
Used to detect the therapeutic range for low molecular weight heparin (LMWH) or unfractionated heparin.
Performing Laboratory
OhioHealth Laboratory Services-RMH Core Coagulation
Specimen Requirements
Draw blood in a light blue-top (citrate) tube(s). Spin down, remove plasma, spin plasma again, and place 1 mL of citrate platelet-poor plasma into a plastic vial. If processed within 4 hours, the specimen does not need to be double spun. Send specimen frozen.
Note: Double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious results.
Specimen Transport Temperature
Frozen/Refrigerate NO/Ambient NO
Only frozen if specimen is not run within 4 hours.
Reference Values
Therapeutic range for LMWH varies with type and manufacturer, but it is typically between 0.4 IU/mL and 1.0 IU/mL. The therapeutic range for unfractionated heparin is 0.3-0.7 IU/mL.
Day(s) Test Set Up
Test performed 24/7, Monday through Sunday
Test Classification and CPT Coding
85520