Test Code LAB90499 Troponin T
Reporting Title
Troponin T
Methodology
Roche Diagnostics, 5th Generation high sensitive cardiac troponin T assay. (hs-cTnT)Electrochemiluminescence Immunoassay
In patients with elevated troponin levels or chest pain of less than 3 hours duration, serial testing is necessary to rule out acute events. The recommended interval for serial testing is 0 (baseline) and 3 hours. A delta troponin value will be included on serial troponin results between 3 and 12 hours apart. In collaboration with the OhioHealth Heart and Vascular Services Clinical Guidance Council, the laboratory has developed algorithms using data from our own patient population. Interpretive comments based on these algorithms will be part of the results.
Performing Laboratory
OhioHealth Laboratory Services-RMH, GMC, DH, DMH, GMH, WECC, PMH, GCMH, BH, OBH, MGH, MH, HMH and SH Core Laboratories
Specimen Requirements
Plasma Only MINT GREEN TUBE
Mint Green (Li-heparin) plasma only. or Li Heparin green top tube no gel. (Hemolyzed specimen is not acceptable.) Spin down and send 1 mL of heparinized plasma refrigerated.
Note: 1. Patients receiving therapy with mouse monoclonal antibodies may have erroneous results.
Due to interference, patients receiving high doses of biotin (>5 mg/day) should wait at least 8 hours after dose before having blood drawn.
Specimen Stability Information
Specimen Type | Temperature | Time |
---|---|---|
Plasma | Refrigerated | 24 hours |
Ambient | ||
Frozen | 12 months |
Specimen Transport Temperature
Refrigerate: OK
Ambient: OK
Frozen NO
Reference Values
Female <6 - 14 ng/L
Male <6 - 22 ng/L
- A comment will be added to Troponin T orders on patients registered as Unknown for gender at the time of testing. The result will flag as critical >14 ng/L. A comment will display for the reference range: Patient sex not reported at the time of testing. Ref Range: Female ≤ 14; Male ≤22.
Delta Troponin T (Abnormal indicates a clinically significant change in troponin concentration)
≤7 ng/L if initial troponin ≤ 35 ng/L
< 20% if initial troponin >35 ng/L
Critical Values
First time troponin value above the reference range
First time abnormal delta troponin
Day(s) Test Set Up
Monday through Sunday; Continuously
Test Classification and CPT Coding
84484